Dryblock Temp. Calibrator TE

The Presys TE Dry Block Temperature Calibrator provides a practical solution for calibrating temperature sensors in industrial plants, maintenance workshops, and calibration laboratories. It generates a controlled temperature inside a metal insert, allowing technicians to compare the readings of thermocouples, resistance temperature detectors (RTDs), and other compatible temperature instruments.

Unlike traditional liquid baths, a dry-block calibrator uses a solid metal insert to transfer heat to the sensor. As a result, technicians can perform routine calibration without filling a bath with calibration fluid. Moreover, selected Presys TE models combine temperature generation with electrical measurement functions for RTDs, thermocouples, milliamps, millivolts, resistance, and temperature switches.

The TE series includes compact low-temperature models and high-temperature models for applications reaching 1200°C. In addition, compatible models support automatic calibration cycles and ISOPLAN® software for recording results and preparing calibration reports. Consequently, the series helps instrumentation teams improve calibration efficiency and maintain organised records.

SKU: TE Serie Categories: ,
Description

Description

Key Features of the Presys TE Series

  • Controlled temperature generation: Produces stable temperatures for checking compatible temperature sensors.
  • Multiple temperature ranges: Offers models for low-temperature calibration and high-temperature applications.
  • Multifunction capability: Selected models combine a dry-block source, reference thermometer functions, and electrical signal measurement.
  • RTD and thermocouple support: Measures compatible sensor signals during calibration.
  • Transmitter signal measurement: Selected models measure 4–20 mA signals and provide a regulated 24 VDC supply for compatible two-wire transmitters.
  • Automatic calibration cycles: Runs programmed temperature points and records results on supported models.
  • Calibration documentation: Connects to compatible ISOPLAN® software for recording results and generating reports.
  • Portable design: Supports field maintenance and laboratory calibration work.

Presys TE Model Range

Model Temperature Range Typical Application
TE-25N −25°C to 155°C Low-temperature sensor calibration
TE-35N −35°C to 155°C Below-ambient and general temperature calibration
TE-50N −50°C to 155°C Lower-temperature calibration
TE-350P Ambient to 350°C General industrial temperature sensors
TE-650P Ambient to 650°C Higher-temperature process instruments
TE-1200P 50°C to 1200°C High-temperature sensors and compatible infrared thermometers

Note: The exact lower and upper limits, accuracy, stability, and available functions depend on the model. Always check the relevant manufacturer’s datasheet before ordering.

How Does the Presys TE Dry Block Calibrator Work?

1. Select the Correct Insert

First, choose a metal insert with a suitable hole diameter for the sensor under test. The correct fit helps improve thermal contact and measurement consistency.

2. Set the Calibration Temperature

Next, enter the required temperature on the calibrator. The instrument heats or cools the insert, depending on the model, until it reaches the selected setpoint.

3. Insert and Connect the Sensor

Place the RTD or thermocouple in the insert. Then connect its electrical terminals to the appropriate measurement input when the model supports direct sensor measurement.

4. Compare the Readings

Once the temperature stabilises, compare the test sensor’s reading with the calibrator’s reference reading or another suitable reference. Repeat the check at the required calibration points.

5. Record the Results

Finally, document the readings and calculate the error at each point. Where supported, automatic calibration and ISOPLAN® software can help technicians record results and generate reports.

Industrial Applications

  • Process instrumentation: Calibrate RTDs and thermocouples used in industrial control loops.
  • Oil and gas: Verify temperature sensors in process equipment, heaters, and utility systems.
  • Power generation: Check compatible temperature instruments used in boilers, turbines, and auxiliary systems.
  • Manufacturing: Verify temperature sensors used in ovens, furnaces, and production equipment.
  • Calibration laboratories: Perform repeatable temperature checks and maintain calibration documentation.
  • Maintenance workshops: Check sensors after installation, repair, or scheduled preventive maintenance.

Benefits of the Presys TE Dry Block Temperature Calibrator

The Presys TE series helps technicians perform routine temperature calibration using a compact, controlled heat source. Furthermore, selected models combine temperature generation and electrical signal measurement, reducing the need for separate instruments.

In addition, programmable calibration cycles can simplify repetitive testing. When technicians use compatible ISOPLAN® software, they can also organise calibration records and prepare reports. Therefore, the TE series supports a more consistent and efficient calibration workflow.

Conclusion

The Presys TE Dry Block Temperature Calibrator offers a flexible range of solutions for industrial temperature sensor calibration. With models covering low-temperature applications through to 1200°C, the series supports a wide variety of maintenance and laboratory requirements. Ultimately, choosing the right model depends on the required temperature range, sensor type, accuracy, and documentation needs.

About Presys

Key Benefits

  • Multifunctional & Integrated Design
  • High Accuracy & Stability
  • Advanced Automation & Documenting
  • Advanced Communication

Products & Features

Presys specializes in a wide range of calibration equipment and software solutions for pressure, temperature, and electrical signals as well as PSV valve workstations. Presys maintains its own laboratory to guarantee traceability to international standards (SI). Many Presys devices include HART protocol support, built-in PDF viewers, and the ability to generate reports directly from the device via Wi-Fi or USB.

  • Temperature Calibrators: dry block and liquid bath, -60 °C to 1200 °C. 
  • Pressure Calibrators: automatic - intrinsically safe - compact
  • Universal Process Calibrators: accuracy up to 0.01% FS - intrinsically safe
  • Mobile Calibration workstations for pressure safety/relief valves (PSV) - compact, transportable, automated digital registration of critical testpoints, 1/2” to 10”, up to 700 bar
  • Smart Connectivity: Compatible with HART, Profibus, USB, Internet, integration with SAP, Azur, Oracle, SQL
  • Calibration Software: calibration automation and management system that centralizes data, schedules calibrations, and issues certificates.